Faults and symptoms during start-up and flow measurement
• Most faults and symptoms occurring with the flowmeters can be eliminated by following the
instructions indicated in the following tables.
• For greater clarity, faults and symptoms in the tables are divided into different groups.
• LED light-emitting diodes on the front panel (status messages)
D
display
I
current output I
P
pulse outputs P and A1
S
status outputs D1, D2, A1 and A2
C
control inputs C1 and C2
Before contacting the KROHNE Service Department,
please read the instructions in the table. THANK YOU.
Group LED
Display
Cause
Remedial action
LED 1
Both LEDs flash
A/D converter range
Reduce flow rate;
exceeded
if unsuccessful, test as
described
in
Section
7.6
Measuring tube drained,-
Fill measuring tube
A/D conv. range exceeded
LED 2
Red LED flashes
Fatal error, hardware and/or
Replace signal converter,
software fault
see Section 8.3
LED 3
Cyclic flashing of red
Hardware fault, watchdog
Replace signal converter,
LED, approx. 1 sec.
trips
see Section 8.3
LED 4
Red LED on
Hardware fault
Replace signal converter,
continuously
see
Section
8.3