05/2003
7 Functional
checks
7.1
Checking the zero with IFC 110 F signal converter, Fct. 3.03
• Set ”zero” flow in the pipeline. Make sure that the measuring tube is completely filled with
liquid.
• Switch on the system and wait 15 minutes.
• Press the following keys for zero measurement:
Key Display
Description
→
If ”YES” is selected in Fct. 3.04 ENTRY CODE,
key in 9-stroke CODE 1 now:
→ → → ↵ ↵ ↵ ↑ ↑ ↑
Fct.
1.00
OPERATION
2x
↑
Fct. 3.00
INSTALL.
→
Fct. 3.01
LANGUAGE
2x
↑
Fct. 3.03
ZERO SET
→
CALIB.
NO
↑
CALIB.
YES
↵
0.00
- - - - - / - - -
Flow rate displayed in set unit, see Fct. 1.04 DISPLAY,
subfunction ”DISP. FLOW”.
Zero is measured, duration approx. 15-90 s. ”WARNING” is
displayed when flow is ”>0”, acknowledge by pressing
↵ key.
STORE
NO
If new value is not to be stored, press
↵ key (3 times)
4 times = return to measuring mode
↑
STORE
YES
↵
Fct. 3.03
ZERO SET
Store new zero value
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode with new zero
7.2
Checking the measuring range Q, Fct. 2.01
• For this test a measuring value can be simulated in the range of -110 to +110 percent of Q
100%
(set full-scale range, see Fct. 1.01 FULL SCALE).
• Switch on the system.
• Press the following keys for checking the measuring range:
Key Display
Description
→
If ”YES” is selected in Fct. 3.04 ENTRY CODE,
key in 9-stroke CODE 1 now:
→ → → ↵ ↵ ↵ ↑ ↑ ↑
Fct.
1.00
OPERATION
↑
Fct. 2.00
TEST
→
Fct. 2.01
TEST Q
→
SURE
NO
↑
SURE
YES
↵
0
PCT.
Current, pulse and status outputs indicate corresponding values.
± 10
PCT.
↑
± 50
PCT.
Select with
↑ key
± 100
PCT.
± 110
PCT.
↵
Fct. 2.01
TEST Q
End of test, actual measured values again available at outputs
(2x) 3x
↵
- - - - - - -
- - - - - / - - -
Measuring mode