5 error messages in measuring mode – KROHNE IFC 110 Converter User Manual
Page 38
Sect. 4.5
Part B IFC 110 F Signal converter
38
IFC 110 F
05/2003
Fct. Text
Description
and
setting
→ LIMIT VAL.
Set limit value for noise/interference suppression
(appears only when ”YES” is selected under ”SPEC. FILT.”, see above)
Setting
range: 01-90 PERCENT of full-scale range Q
100%
see Fct. 3.02, subfunction ”FULL SCALE”
Press
↵ key to change to subfunction ”LIMIT CNT.”.
→ LIMIT CNT.
Totalizer active when exceeding limit value (see ”LIMIT VAL.” above)
(appears only when ”YES” is selected under ”SPEC. FILT.”)
Setting
range: 001-250
Press
↵ key to return to Fct. 3.06 ”APPLICAT.”.
3.07
HARDWARE
Determine HARDWARE functions
→ TERM.A1
Terminal A1
● PULSOUTP.
● STATUSOUTP.
Select with key
↑.
Press key
↵ to transfer to subfunction ”SELFCHECK”.
→ SELFCHECK
Carry out self check? See Section 5.18.
● YES
● NO
(testing different parameters)
Press key
↵ to transfer to subfunction “FIELD CURRENT”.
→ FIELDCUR.
Determine field current
● INTERNAL
● EXTERNAL (only with power driver, see Sect. 8.6)
Press
↵ key to return to Fct. 3.07 ”HARDWARE”.
4.5
Error messages in measuring mode
The following list contains all errors which may occur during flow measurment. Errors are displayed when
”YES” is selected in Fct. 1.04 DISPLAY, subfunction ”DISP. MSG.”.
Error message
Description of error
Elimination of error
LINE INT.
Power failure Note:
Cancel error in RESET/QUIT. menu
no counting during power failure
Reset totalizer if necessary.
OVERFLOW I
Current output overranged.
Check instrument parameters and
or
(flow rate > measuring range)
correct if necessary. After elimination of
OVERFL. I2
the cause, the error message is
cancelled
automatically.
See Sections 6.4 and 6.7.
OVERFLOW P
Pulse output P
Check instrument parameters and
or
or
correct if necessary. After elimination of
OVERFL. P2
Pulse output range P2 exceeded
the cause, the error message is
(flow rate > modulation range)
cancelled automatically.
See Sections 6.4 and 6.7.
I SHORT or
*
Current output I or I2 externally
Check mA loop and increase load using
I2 SHORT
shorted or load < 15
Ω
additional resistor if necessary.
I OPEN or
*
mA loop interrupted by current
Check mA loop and reduce load to
I2 OPEN
output I or I2 or load > 500
Ω 500
Ω if necessary.
TOTALIZER
Overflow of internal totalizer
Delete error message in RESET/QUIT
menu, see Sect. 4.6
ADC
Analog/digital converter range
Set Fct. 3.06, subfunction ADC GAIN
exceeded
to ”10”. See Sections 6.4 and 6.7.
If error message does not disappear,
consult
factory.
ADC-PARAM.
Check sum error
Replace ADC printed circuit board
ADC-HARDW.
Hardware error A/D converter
Replace ADC printed circuit board
ADC GAIN
Hardware error A/D converter
Replace ADC printed circuit board
FC-HARDW.
Hardware error on field current PCB
Replace field current PCB
FATAL.ERROR
Fatal error, measurement interrupted
Replace electronic unit or consult factory.
EP PARAM.
Parameters of “Empty Pipe” are wrong
Error shut off automatically EPD function.
Please check calibration values!
Value Empty Pipe – Value Full pipe
≥ 10
Values must be in range of 0 – 150.
*
only for active operation