beautypg.com

KROHNE IFC 110 Converter User Manual

Page 53

background image

Part C Special applications, functional checks, service, and order numbers Sect. 6.1

05/2003

IFC 110 F

53

5.19 Limit

switches


Fct. 1.07 - 1.10 Status outputs A1, A2, D1 or D2
(Define operating mode of output terminals A1, see Section 5.18)
Press

key

Press

key as often as required to set one of the status outputs to ”TRIP. POINT”


Press

key to change to ”Characteristic” (flow direction).


Select:

+ DIR.

– DIR.

Select with

↑ or ↓ keys

2 DIR.

}


Press

key to change to numerical setting, 1st number (cursor) flashes.

Change flashing number (cursor) with keys

and ↓. Useandkeys to shift cursor

1 place to right or left.

Display: XXX – YYY


Setting ranges: XXX value = 0 – 150% of Q

100%

YYY value = 0 – 150% of Q

100%

hysteresis

≥ 1% (= difference between XXX and YYY values)


Switching behaviour (NO/NC contact) and hysteresis are adjustable.


NC contact XXX value > YYY value

NC contact XXX value < YYY value

Switch closes when flow exceeds XXX value

Switch opens when flow exceeds YYY value

Example: XXX = 55%

Example: XXX = 45%

YYY = 45%

YYY = 55%

hysteresis = 10%

hysteresis = 10%


Please note:

if two status outputs (e.g. D1 and D2) are activated

it is possible to have e.g. min. and max. values signalled.