KROHNE IFC 110 Converter User Manual

Page 36

background image

Sect. 4.4

Part B IFC 110 F Signal converter

36

IFC 110 F

05/2003

Fct. Text

Description

and

setting

3.0

INSTALL.

Installation menu

3.01

LANGUAGE

Select language for display texts

● GB / USA (English)

● S (Swedish)

● D (German)

● other languages on request

● F (French)

Press

key to return to Fct. 3.01 “LANGUAGE”.

3.02

FLOWMETER

Set data for flow sensor

DIAMETER

Select size from meter size table

● DN 2.5 - 1200 mm equivalent to 1/10 - 48 inch

● DN 1300 - 3000 mm equivalent to 52 - 120 inch (see Sect. 8.6)

Select with

key.

Press

key to change to subfunction “FULL SCALE”.

FULL SCALE Full-scale range for flow Q

100%

To set, refer to Fct. 1.01 “FULL SCALE”.

Press

key to change to subfunction “GK VALUE”.

VALUE P

Pulse value for pulse output P (Fct. 1.06 ”VALUE P”) and/or

and/or

for the 2nd pulse output A1 (Fct. 1.07 ”VALUE P2”) has been changed.

VALUE P

With the ”old” pulse values the output frequency (F) would have been exceeded
or would not have been reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new value.

GK VALUE

Set primary constant GK

See instrument nameplate of flow sensor.

Range:

● 1.0000 - 15.000

Press

key to change to subfunction “FIELD. FREQ.”.

FIELD FREQ. Magnetic field frequency

Values: 1/2, 1/6, 1/18 and 1/36 of power frequency, see instr. nameplate.

Press

key to change to subfunction “FLOW DIR.”;

on DC instruments change to subfunction “LINE FREQ.”.

LINE FREQ.

Power frequency customary in the country where the instrument is used

Please

note: this function is limited to instruments with DC supply unit (24 V DC)

to suppress line frequency interferences.

Values:

50 Hz and 60 Hz

Press

key change to subfunction “FLOW DIR.”.

FLOW DIR.

Define flow direction (in F/R mode: forward flow).

Set according to direction of arrow on flow sensor:

● + DIR.

● – DIR. Select using

key.

Press

key to return to Fct. 3.02 “FLOWMETER”.

3.03

ZERO SET

Zero calibration

Note: carry out only at ”0” flow and with completely filled measuring tube!

Precautionary

query

● CALIB. NO

Press

key to return to Fct. 3.3 “ZERO SET”.

● CALIB. YES

Press

key to start calibration.

Duration approx. 15-90 s (depending on magnetic field
frequency), current flow rate displayed in the selected unit
(s. Fct. 1.04 ”DISP. FLOW”).

A ”WARNING” sign appears when flow rate ”>0”;

acknowledge by pressing

key.

● STORE NO

(do not store new zero value)

● STORE YES

(store new zero value)

Press

key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?

● NO

(= entry with

→ only)

● YES (= entry with → and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )

Press

to return to Fct. 3.04 “ENTRY CODE”.