beautypg.com

KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 91

background image

91

8.8

Checking the Primary Head


Remember! Always switch the power off before opening the converter housing.

Required tools and test equipment
– Phillips screwdriver
– Multimeter
– Special wrench to unscrew converter housing lids

8.8.1 Compact Meter

A. Screws that hold the display
PCB in place.

B. Ribbon cable connecting display
PCB to converter electronics.

C. 10-pin sensor connector.

D. Screws that hold the converter
electronics in place.

P. Power supply circuit board

Preliminary preparation
– Remove front lid of electronic housing
– Unscrew the two screws ”A” that hold down the display PCB and fold the ribbon cable and

PCB carefully to one side.

– Unplug blue primary connector ”C” from amplifier board

10-Pin Sensor Connector ”C” (Connector between primary and converter)

(g

re

y)

(w

h

it

e

)

(b

lue)

(g

reen)

(v

io

le

t)

(b

la

ck

)

(b

ro

w

n

)

(o

range)

(y

e

llo

w

)

(r

ed)

Colours in parenthesis are valid for Ex-converters

Testing the resistance values of
the sensors and exciter

Typical values

Assessment of
measured values

1

Measuring tube driver:
Measure between grey and black

30 - 50 ohm

2

To check sensor A and B:
Measure between:
green and violet (Sensor A)
white and yellow (Sensor B)

50 - 130 ohm

Measurements outside
typical values:
Primary faulty
Replace or call Krohne
service.

3

To check the temperature
sensor (RTD)
:
Measure between blue and red.

500 - 550 ohm

(ambient temperature

dependent)

Measurements inside
typical values:
Sensor OK

4

To check the strain gauge:
Measure between orange and red.

400 - 600 ohm

C