Low-flow cut-off

0 - 10% of nominal full-scale range

Time constant for flow

0.5 - 20 seconds (optionally: 0.2 - 20 seconds)

Power supply

Standard

230 V AC

±

10%

200 V AC

±

10% 48 - 63 Hz

115 V AC

±

10%

100 V AC

±

10%

Special versions

21, 24, 42, 48 V AC, +10/-15%, 48 - 63 Hz

24 V DC,

±

30%

Power consumption

AC : 18 VA

DC : 10 W

Operator control / interfaces

Keypad

3 keys

Local display

Type

3-line, illuminated LCD display

1st (top) line: 8 character, 7 segments for numerals and signs

2nd (middle) line: 10-character, 14 segments for texts

3rd (bottom) line: 6 markers

for status identification

Function

actual measured value, forward, reverse or sum totalizer
(7 characters), each can be set for continuous or cyclic
display, and status output

Meas quantities. and

units

Plain text language

see page 89 “Measured quantities and units”

English, German French

Magnetic sensors MP

same function as the 3 keys, operation by means of hand-held
bar magnet without opening the housing

Communications options

RS 232 adaptor and Config. Software

Operating via a MSDos PC. For further information see
Product Guide 6 “Communications techniques”.

HART-System

Via handheld communicator. For further information see
Product Guide 6 “Communications techniques”.

RS 485/ Modbus Serial communications

For detailed protocol information please contact Krohne

Housing material

die-cast aluminium with polyurethane finish