beautypg.com

KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 67

background image

67

CF1 to CF4 can be found from menus Fct. 3.9.1 to 3.9.4 To find the remaining data proceed as
follows.


Key

Line 1

Line 2

Notes:

Fct. 3.9.(10)

D. REF. HIGH

(MEAS.
VALUE)

(SET VALUE)

210.1234

HZ

Record this frequency value.

22.1

°C

Record this temperature value.

467.05

STRAIN

Record this strain value.


or

1200.1

(WATER)


(kg) / min

If WATER (or AIR for menu 3.9.6)
was selected then this will be
displayed here. If a fluid other than air
or water was selected then the
density input by the customer will be
displayed. If required these values
can be edited here. Note if the value
is edited CF1 and CF2 will be
adjusted.

Fct. 3.9.(10)

D. REF. HIGH

Fct. 3.9.(11)

D. REF. LOW

Repeat for the LOW calibration point.


When reviewing the calibration data the frequency, temperature and strain values cannot
normally be edited. If a replacement converter is being fitted then it may be necessary to copy
the calibration points into the new set of electronics. (CF1 to CF5 must be copied in all cases).
Frequency , temperature and strain data can be edited if the operator first enters the
PARAM.CODE.4 password in menu 3.3.8 (see section 5.14.4).