KROHNE CORIMASS G-MFM 4085 EN User Manual
Page 18
18
2.2
Connection to power
–
Note information given on the instrument data plate (voltage, frequency)!
–
Electrical connection in conformity with IEC 364 or equivalent national standard.
Special regulations apply to installation in hazardous areas. Please refer to separate "Ex"
installation instructions.
– The PE protective ground conductor must be connected to the separate U-clamp
terminal in the terminal box of the signal converter.
–
Do not cross or loop the cables in the terminal box of the signal converter. Use separate
(PG or NPT) cable glands for power and output cables.
–
Ensure that the screw thread of the round cover on the terminal box is well greased at all
times.
NOTE: The grease used must be non-corrosive to aluminium; typically it must be resin- and
acid-free.
– Protect sealing ring from damage.
Outputs and
connections
see Sect. 2.3
Power and signal connections for MFC 085 K
2.3
Inputs and outputs
The table below shows the input/output connection for the converter. The exact configuration
depends on which optional output modules were fitted in the factory.
Table of input/output connections
Trm.
No.
Option 1
(Current, pulse, status
output and control input )
Option 2
*
(2 current,NGI
status output)
5
Common (-)
Common (-)
6
Current output (+)
Current output 1 (+)
4
Control input
Control input.
4.1
Pulse output
Current output 2 (+)
4.2
Status output (active)
Status output (passive)
*
The inputs/outputs share a common signal ground that is galvanically isolated from ground
(PE).
Please ensure that the information about power given on the
data plate corresponds to the locally available mains voltage.
PE
5
6
4
4.1
4.2
11
12
N L AC
–
+ DC