Process alarm characteristics

For example, a process requires that the process fluid temperature stays within the range 30 to
40°C and the low level signal is required to indicate that the temperature has gone out of range.
Begin from measuring mode

Return to measuring mode.
NOTE:
For functions other than range checking Fct. 3.5.3. and 3.5.4. are not accessible.

Key

Display
line 1

line 2

→↑↑

Fct. (3).0.

INSTALL

4x

Fct. 3.(5).0.

ALARM.OUT.A

Fct. 3.5.(1).

FUNCTION A

(OFF)

(MASS FLOW)

:

(MASS TOTAL)

:

(DENSITY)

(TEMPERATURE)

Fct. 3.5.(1).

FUNCTION A

Fct. 3.5.(2).

ACTIV.LEVEL

(ACTIVE.HIGH)

(ACTIVE LOW)

Fct. 3.5.(2).

ACTIV.LEVEL

Fct. 3.5.(3).

MIN. LIMIT

Input minimum temperature

Fct. 3.5.(3).

MIN. LIMIT

Fct. 3.5.(4).

MAX. LIMIT

Input maximum temperature

Fct. 3.5.(4).

MAX. LIMIT

4x