Status messages
ERROR
MESSAGES
TYPE
COMMENT
SAMPLING
Severe
PLL out of range
SENSOR A
Severe
Sensor A voltage signal less than 5% of desired value
SENSOR B
Severe
Sensor B voltage signal less than 5% of desired value
RATIO A/B
Severe
One sensor signal much larger than the other
EEPROM
FATAL
Unable to save data in EEPROM. Hardware fault.
SYSTEM
FATAL
Indicates software error, will always occur with
WATCHDOG
WATCHDOG
Severe
Reset due to SYSTEM error or temporary power supply
drop-off
NVRAM
Severe
NVRAM check sum error, previous data lost
DC A
Severe
DC voltage part of sensor A is larger than 20% of ADC max
DC B
Severe
DC voltage part of sensor B is larger than 20% of ADC max
NVRAM FULL
Light
NVRAM has exceeded its specified number of write cycles
MASS FLOW
Light
Mass flow rate > 2 nominal flow *
ZERO ERROR
Light
Mass flow rate at zero adjust is larger than > 20% of
nominal (100%) flow rate *
TEMPERATUR
Light
Temperature > outside operating range
STRAIN
Light
Strain out of operating range
CURRENT.SAT
Output
Current output saturated **
FREQ.SAT
Output
Frequency output saturated **
ALARM.OUT.A
Output
Process alarm limit check exceeded **
ROM DEF
Light
EEPROM check sum error, defaults loaded from ROM
TOTAL O/F
Light
Custody transfer only. Mass total has overflowed the
display, i.e. it has gone from 99999999 to 00000000
TEMP.CUST
Light
Custody transfer only. Operating temperature has drifted
by more than
±
30
°
C from the zero calibration temperature
POWER.FAIL
Light
Custody transfer only. There has been an interruption of
power to the converter.
*
Actual mass flow rate is too big or manual zero offset PUTIN.VAL in Fct. 1.1.1 was
programmed incorrectly.
**
Change output range to avoid saturation.