40
Fct. No.
Text
Description and settings
3.9.0
TUBE PARAMS
Submenu 3.9.0 primary head specific parameters
Password protected, see Fct. 3.8.8
3.9.1
CF1
Density coefficient 1
Input the value stamped on the primary head’s data plate
or perform on site calibration as described in Sect. 5.12.
3.9.2
CF2
Density coefficient 2
Input the value stamped on the primary head’s data plate
or perform on site calibration as described in Sect. 5.12.
3.9.3
CF3
* Reference strain
Displays the value stamped on the primary head’s data
plate.
3.9.4
CF4
* Reference temperature
Displays the value stamped on the primary head’s data
plate.
3.9.5
CF5
*Primary head constant
Displays the value stamped on the primary head’s data
plate.
3.9.6
DSS CF6
Density strain slope
Displays the value which is mentioned on the cal. Sheet
3.9.7
DTS CF 7
Density temperature slope
Displays the value which is mentioned on the cal. Sheet
3.9.8
FSS CF8
Slope strain
Displays the value which is mentioned on the cal. Sheet
3.9.9
FTS CF9
Slope temperature
Displays the value which is mentioned on the cal. Sheet
3.9.10
D.REF.HIGH
Density adjustment, High Point
3.9.11
D.REF.LOW
Density adjustment, Low Point
3.10.0
CONC.MEAS
Concentration measurement option when installed
3.10.1
SOLUTE R20
See separate Concentration instruction manual
3.10.2
SOLUTE K1
See separate Concentration instruction manual
3.10.3
SOLUTE K2
See separate Concentration instruction manual
3.10.4
LIQUID
See separate Concentration instruction manual
3.10.5
LIQUIDR20
See separate Concentration instruction manual
3.10.6
LIQUID K1
See separate Concentration instruction manual
3.10.7
LIQUID K2
See separate Concentration instruction manual
3.11.0
SERIAL I/O
RS485 or Modbus option, if installed
3.11.1
PROTOCOL
See separate RS485 or Modbus instruction manual
3.11.2
ADDRESS
See separate RS485 or Modbus instruction manual
3.11.3
BAUDRATE
As per 3.11.1
* Only accessible if allowed through password in Fct. 3.8.8.