beautypg.com

KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 74

background image

74

5.14.4 Primary head type and tube parameters (CF1 - 5)

The primary head type and tube parameters are factory set and should not normally be changed
by the customer. They are only normally needed if the converter has to be replaced in the field.
In this case, the new converter must be programmed with the correct primary head type, and
parameters, CF1 to CF5, must be entered as printed on the data plate.
In order to prevent accidental changes to key parameters CF3 and CF5 and the Primary head
type, an additional level of password protection is included. The operator may freely look at
these parameters as ”READ ONLY” but he may only be able to change it when he has first
entered the correct PARAM.CODE 4 password.

Fct. (1).0

OPERATION

↑↑

Fct. (3).0

INSTALL

→↑

×

7

Fct. 3.(8).0

USER DATA

→↑

×

7

Fct. 3.8.(8)

PARAM.CODE.4

CodE 4

– –

↵↑

OK

Fct. 3.8.(8)

PARAM.CODE.4

The operator may now edit, (if required), CF3, CF4, CF5 as well as PRIMRY.TYPE. Having
made the changes the operator should save them and return to the main displays. Once back
in the main displays any subsequent attempts to edit these parameters will be blocked until the
PARAM.CODE.4 password is re-entered.

To edit these parameters:

Key

Display
line 1

line 2

Fct. 3.8.(8)

PARAM.CODE.4

↵↵

Fct. (3).0

INSTALL

→→

4x

Fct. 3.1.(5)

PRIMRY.TYPE

(10G)

T

(100G)

T

(300G)

T

(300G)

T

(300G)

T+

Select the correct size and type (T,
T+, Z, Z+) as printed on the data
plate.

Fct. 3.1.(5)

PRIMRY.TYPE

Fct. 3.1.(6)

CF5

(0)16.000

CF5

Enter CF5 as printed on the data
plate.

Fct. 3.1.(6)

CF5

Fct. 3.1.0.

INSTALL

8x

Fct. 3.9.0.

TUBE PARAM

Fct. 3.9.1.

CF1

Enter CF1 to CF4 (Fct. 3.9.1. to
3.9.4.) as printed on the data plate.

4x