KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 33

background image

33

Fct. No.

Text

Description and settings

3.0

INSTALL.

Main menu 3.0 Installation

3.1.0

BASIS.PARAM

Submenu 3.1.0 Base data

3.1.1

ZERO SET

Zero adjustment.
Use the

key to select between MEASURE.VAL.

and SET VALUE then press the

key.

* MEAS.VALUE (ensure "ZERO" flow in the pipeline)
1) Select: CALIB.YES or NO
2) If YES: Calibration (approx. 20 sec. duration)

Display: Actual flow rate as percent of the
maximum rated flow for the primary head. (Q

100%

)

3) Select: ACCEPT YES or NO
* SET.VALUE Direct input of a zero flow offset.
Units: As selected by Fct. 1.2.1 or 3.2.1

3.1.2

L.F. CUTOFF

Low flow cutoff
Value: 0 to 10 percent of nominal flow

3.1.3

TIME CONST.

Time constant for output of measured values
Range 0.5 ... 20 sec. (Option: 0,2 ... 20 sec.)

3.1.4

STANDBY

Use the key

to switch between three modes

of operation, then press

:

*

MEASURE

*

STANDBY (tube vibrating, Mass Flow set to zero)

*

STOP (tube drive stopped)

Note:

It is not possible to switch directly
from STOP to STANDBY.

3.1.5

PRIMRY.TYPE

Type of the primary head **
Using the

key select the primary head type

that is connected to the converter:
* 10 G

* 800 G

*100 G

*1500 G

*300 G

*3000 G

Then press

key to select material field, then

key to

select:
* T * T+ * Z * Z+

as per the data plate

3.1.6

CF5

Primary head constant. **
Displays the primary head constant as stamped
on the primary head’s data plate. (Password protected)

3.1.7

FLOW DIR.

Define direction of flow.
Select either FORWARD or BACKWARD

3.1.8

FLOW MODE

Define whether bi-directional or uni-directional
flow is expected.
Select either:
* FLOW > 0 (Ignore negative flows)
* FLOW < 0 (Ignore positive flows)
* FLOW +/- (Allow positive and negative flows)

**

These menus are protected by the Code 4 password, see Fct. 3.8.8