Fct. No.

Text

Description and settings

3.3.0

CUR.OUTP. I

Submenu Current output I

For systems with 2 or more current outputs see Sect. 4.7

3.3.1

FUNCTION I

Function current output I

* OFF (O/P current = 0 mA)

* MASS FLOW (Mass flow in range MIN [Fct. 3.3.3] to

MAX [Fct. 3.3.4] output as current in range [Fct 3.3.2]

0/4-20mA)

* DENSITY (Density in range MIN [Fct. 3.3.3] to MAX

[Fct. 3.3.4] output as current in range [Fct 3.3.2]

0/4-20 mA)

* TEMPERATUR (Temperature in range MIN [Fct. 3.3.3]

to MAX [Fct. 3.3.4] output as current in range

[Fct 3.3.2] 0/4-20 mA)

* VOL.FLOW (Volume flow in range MIN [Fct. 3.3.3]

to MAX [Fct. 3.3.4] output as current in range

[Fct 3.3.2] 0/4-20 mA)

SOLUTE.FLOW

Concentration measurement

CONC. BY MASS

functions available if installed

CONC.BY

(see sep. instruction manual).

* DIRECTION (Negative flow gives current of 0/4 mA,

positive flow gives current of 20 mA)

3.3.2

RANGE I

Range for current output I: Select from the following

by pressing

key and then

key

* 0-20 mA

* 4-20 mA

* 0-20/22 mA (O/P = 22 mA when error detected)

* 2/4-20 mA (O/P = 2 mA when error detected)

* 3.5/4-20 mA (O/P = 3.5 mA when error detected)

3.3.3

MIN.VALUE

Value of measured quantity as set by Fct. 3.3.1

or

MIN. FLOW,

that corresponds to the minimum output current

or

MIN. DENSITY

(0 or 4 mA as set by 3.3.2)

or

MIN. TEMP.

or

MIN V.FLOW

or

MIN.CONC.

Menu not available if Function 3.3.1 is set to OFF or

DIRECTION

3.3.4

MAX.VALUE

Value of measured quantity as set by Fct. 3.3.1

or

MAX. FLOW,

that corresponds to an output current of 20 mA

or

MAX. DENSITY,

or

MAX TEMP

or

MAX V.FLOW

or

MAX.CONC

Menu not available if Function 3.3.1 is set to OFF or

DIRECTION