For systems with concentration software their specific extra displays will be added to the above
list from Fct. 1.2.9 onwards.

The displayed measurement function can be changed in the measurement mode by pressing

key.

The setting of the display output is demonstrated for the example of displaying the mass flow
rate in kg/h.

Starting from the display of measurement values, the following programming steps are
necessary.


Begin from measuring mode.


After pressing

key the display shows :


0000.0000 (kg)/min


This display means that mass flow will be displayed in units of kg/min to an accuracy of 4
decimal places.

The brackets around 'kg' indicate the cursor position, these characters will be flashing on the
display. The flashing value can now be changed with the

key. By pressing the

key the

cursor moves to 'min' which starts to flash.

This dimension may now also be changed by using the

key. After pressing the

key again,

the cursor moves to the output format of the numeric value which can then be edited.


To change the display to kg/h with 5 decimal places, follow the following procedure.


Key

Display

line 1

line2


Fct. (1).0

OPERATOR

2x

Fct. (3).0.

INSTALL

Fct. 3.(1).0.

BASE DATA

Fct. 3.(2).0.

DISPLAY

→↑↑

Fct. 3.2.(3).

MASS FLOW


Key

Display

line 1

line 2


0000.0000

(kg)/min

0000.0000

kg/(min)

0000.0000

kg/(h)

0000(.)0000

kg/h

00000(.)000

kg/h

000000(.)00

kg/h

0000000(.)0

kg/h

00000000(.)

kg/h

0(.)0000000

kg/h

00(.)000000

kg/h

000(.)00000

kg/h

Fct. 3.2.(3).

MASS FLOW