KROHNE CORIMASS G-MFM 4085 EN User Manual
Page 41
41
Fct. No.
Text
Description and settings
4.0
PARAM.ERROR
Main Menu 4.0 Parameter error
4.1
Not Used
4.2.0
CUR.OUTP.I
Range settings incorrect
LOW SCALE
≥
FULL SCALE
4.2.1
LOW SCALE
Low scale range for current output I see Fct. 3.3.3
4.2.2
FULL SCALE
Full scale range for current output I see Fct. 3.3.4
4.3.0
ZERO
Zero calibration incorrect.
The measured zero offset must be less than ±10 %
of the primary head’s full scale flow rating.
4.3.1
ZERO SET
Zero calibration see Fct. 3.1.7
4.3.2
PRIMRY.TYPE
Type of primary head see Fct. 3.1.5
4.4
Not Used
4.5
Not Used
4.6.0
PULS.OUTP. P
Range setting incorrect
LOW SCALE
≥
FULL SCALE
4.6.1
LOW SCALE
Low scale range for pulse output see Fct. 3.4.3
4.6.2
FULL SCALE
Full scale range for pulse output see Fct. 3.4.4
4.7.0
PROC. ALARM
Minimum and maximum limits incorrect
MIN.LIMIT > 96 % of MAX.LIMIT
4.7.1
MIN.LIMIT
Minimum limit for range checking see Fct. 3.5.3
4.7.2
MAX.LIMIT
Maximum limit for range checking see Fct. 3.5.4
4.8.0
SYS.CTRL.S
Minimum and maximum limits incorrect
MIN.LIMIT > 96 % of MAX.LIMIT
4.8.1
MIN. LIMIT
Minimum limit for condition checking see Fct. 3.7.3
4.8.2
MAX. LIMIT
Maximum limit for condition checking see Fct. 3.7.4
4.5
Reset / Quit Menu - Totalizer reset and status indication acknowledgement
Totalizer reset
Button
Display
Description
10.36
kg
Measurement mode
↵
CodE 2
––
Enter access Code 2 for reset/quit menu:
↑
→
↑
→
RESET.TOTAL
Totalizer reset menu
Only appears if ”yes” programmed in Fct. 3.8.5. Reset
enable No or Yes. If ”no” is programmed ”status light”
only appears. See next section.
→
RESET.YES
If the reset function is enabled RESET YES will be
shown, press
↵
to execute the function.
To cancel the reset operation press
↑
to get RESET NO
and then press
↵
If the reset function is disabled by menus Fct. 3.8.5 or
3.8.6 then BLOCKED is displayed. Press
↵
to continue
↵↵
0.00
kg
Assuming RESET YES was selected the totalisers will
now be cleared.