beautypg.com

KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 32

background image

32

Fct. No.

Text

Description and settings

2.0

TEST

Main Menu 2.0. Test functions

2.1

TEST DISP.

Carry out display test.
Start with the key

(Duration of test approx. 30 sec.).

Stop test at any time with the

key.

2.2

TEST I

Test current output I
* SURE (NO). Use the

key to select YES, then press

.

* 0 mA will be output from the converter. Use the

key

to select test currents from the list below.

0 mA, 2 mA, 4 mA, 10 mA, 16mA, 20 mA, 22 mA.
To exit test mode,

press the

key at any time.

2.3

TEST P

Test frequency output P
* SURE (NO). Use the

key to select YES ,then

press

key.

2.3.1

FREQUENCY

* LEVEL LOW 0 volt DC level will be output from the

converter.
Use the

key to select test signals from the list below.

* LEVEL HIGH (+ V volts dc)
* 1 Hz

* 100 Hz

* 10 Hz

* 1000 Hz

2.3.2

TEST PULSE

* Test Pulse
Use the

key to select desired pulse width from the list

below:

0.4 mSec

100.0 mSec

1.0 mSec

500.0 mSec

10.0 mSec

Then press

. The system now sends pulses of the

required width. To stop the test press

twice.

2.4

TEST A

Test alarm output
* SURE (NO). Use the

key to select YES, then

press

* LEVEL LOW. 0 Volts is output on the alarm terminal.

Press the

key to switch output to:

* LEVEL HIGH . +24V dc is output on the alarm terminal.

To exit test mode, press the

key at any time.

2.5

TEST INP.E

Test control input
The actual input level, HI or LO, and the selected
functions are displayed see Fct. 3.6.1
End test by pressing the

key.

2.6

TEST TEMP.

Test temperature and strain gauge
Start with the

key. The temperature in °C is displayed.

Use the

key to display the temperature in °F. Use the

key again to display strain. End the test by pressing

key

2.7.0

TEST.PRIMRY.

Sub menu 2.7.0 Test primary head values.

2.7.1

SENSOR A

Monitor the amplitudes of sensor A and B

2.7.2

SENSOR B

as percentage of their max. value. (80% is ideal)
Start test with the

key. End the test with the

key.

2.7.3

FREQUENCY

Monitor the primary head frequency.
Start test with the

key. End the test with the

key.

2.7.4

INSTAL.FACT.

Monitor the primary head's drive level.
Start test with the

key. End test with the

key.