Current output switched

off.

Switch on as per Fct.

3.3.1

I

2

Wrong display on

connected instrument

Current programming

not correct.

Correct the programming

as per Fct. 3.3.1 - 3.3.4

I

3

Fluctuating display on

connected instrument

Time constant too small

Increase the time

constant as per Fct. 3.1.3

Group P

P1

Connected totalizer not

counting

Connection / polarity not

correct.

Check and correct as per

Section 2.3

Totalizer or external

supply voltage faulty.

Check output with

totalizer:

Test OK

Check cabling and totalizer.
Check external voltage supply

Test faulty

Pulse output faulty. Replace
converter or call Krohne
Service.

Alarm output is used as

external voltage supply, a
possible electrical short-circuit
or alarm/pulse output faulty.

Check connections as per

Section 2.3. Voltage between
terminals 5 and 4.2 approx. 24
V. Correct short circuit if
present. If still faulty the
alarm or pulse output is faulty.
Re-place converter or call
Krohne Service.

Pulse output is switched

off

Switch on as per Fct.

3.4.1

P2

Fluctuating pulse rate

Time constant too small

Increase time constant as

per Fct. 3.1.3

P3

Pulse rate too high or too

low

Programming of pulse

output incorrect.

Correct programming as

per Fct. 3.4.1 - 3.4.4

External induced noise due

to low quality cable or
unscreened cable.

Check cabling and replace

with screened cable.

See Section 2.3

Group A

A1

Alarm output not

functioning

Connection / polarity

incorrect.

Correct as per Sect. 2.3

Alarm output or external

instrument faulty

Program alarm output to

”direction” as per Fct. 3.5.1.
Set flow direction to negative
and check alarm output.

Test OK

Check external instrument
and if necessary replace.

Test faulty

Alarm output faulty. Replace
converter or call Krohne
Service.

Alarm output switched off

Switch on as per Fct.

3.5.1

A2

Incorrect voltage level at

output terminals (Hi/Lo)

Incorrectly programmed in

Fct. 3.5.2

Correct as follows:

Hi = 24 V

Lo = 0 V