Additional input/output options
Tr
m.
No.
Option 4*
(Current and
RS485)
Option 5*
(Current and
Modbus)
Option 6
(1 Current,
1 Dual
phase pulse
output and
control
input)
Option C
(2 current,
pulse and
control
input)
Option D
(3 current
and
pulse)
Option E
(3 current
and
control
input)
Option F
(3 current
and
status
output)
5
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
6
Current
output 1 (+)
Current
output 1 (+)
Current
output 1 (+)
Current
output (+)
Current
output 1 (+)
Current
output 1 (+)
Current
output 1 (+)
4
TX/RX
TX/RX
Control Input
Current
output 2 (+)
Current
output 2 (+)
Current
output 2 (+)
Current
output 2 (+)
4.1
TX/RX
TX/RX
Pulse
Output A
Control Input
Current
output 3 (+)
Current
output 3 (+)
Current
output 3 (+)
4.2
+5V
+5V
Pulse
Output B
Pulse
Output
Pulse output
Control Input
status output
(passive)
The pulse out alarm outputs where selected are passive.
*
Refer to separate RS 485 or Modbus manual
2.4
Connection of remote meters
The G meter can be supplied as remote meter with 5 m cable. Under no circumstances should
the cable be cut shorter or joined to increase its length. The meter is calibrated with this 5m
length. Any changes will influence the performance of the meter.
There are two different configurations of remote meter, on the first version the cable is fixed at
the converter end and the second has a terminal junction box at the converter end.
115