KROHNE CORIMASS G-MFM 4085 EN User Manual
Page 37
37
Fct. No.
Text
Description and settings
3.5.0
ALARM.OUT.A
Sub menu 3.5.0 Process alarm output
3.5.1
FUNCTION A
Function for alarm output P
* OFF (Output goes to its inactive state)
* MASS FLOW (Alarm active if mass flow goes outside
limits as set in Fcts. 3.5.3 and 3.5.4)
* MASS TOTAL (Alarm active if totaliser goes outside
limits as set in Fcts. 3.5.3 and 3.5.4)
* DENSITY (Alarm active if density goes outside limits
as set in Fcts. 3.5.3 and 3.5.4)
* TEMPERAT. (Alarm active if temperature goes outside
limits as set in Fcts. 3.5.3 and 3.5.4)
* VOLUME.FLOW (Alarm active if volume flow go
outside limits as set in Fcts. 3.5.3 and 3.5.4)
* VOL.TOTAL (Alarm active ...
Solute flow
Concentration option
Conc. by mass
if installed. See separate
Conc. by volume
instruction manual
*
I
1.SAT (Alarm active if value output on current output
exceeds the range as set in Fct. 3.3.3 and 3.3.4)
* P 1.SAT (Alarm active if value output on pulse output
is either:
> 1.3 x Max Limit as set in Fct 3.3.4 or
< Min Limit as set in Fct 3.3.3
* ANY O/P.SAT (Alarm active if value output on either
current or pulse output exceeds the selected ranges)
* SEVERE ERR. (Output active if a severe error is
detected)
* ALL MSG. (Output active if any warnings occur)
* DIRECTION (Output active for positive flows, inactive
for negative flows)
3.5.2
ACTIV.LEVEL
Select the desired voltage level for the active state
* ACTIVE.HIGH (24 V dc)
* ACTIVE LOW (0 V dc)
3.5.3
MIN. LIMIT
Minimum allowable value for functions
TOTAL MASS, MASS FLOW, DENSITY,
TEMPERATUR and VOLUME.FLOW
Units: depend on function but will correspond
to those set in Fcts. 3.2.1 to 3.2.5
or
Not accessible for all other functions
3.5.4
MAX. LIMIT.
Maximum allowable value for functions
MASS TOTAL, MASS FLOW, DENSITY,
TEMPERATUR and VOLUME.FLOW
Units: depend on function but will correspond
to those set in Fcts. 3.2.1 and 3.2.5
or
Not accessible for all other functions
}