38
Fct. No.
Text
Description and settings
3.6.0
CTRL.INP.E
Submenu 3.6.0 Control input
3.6.1
FUNCTION E
Function of the control input
* OFF (control input inactive)
* STANDBY (When active converter switches to
STANDBY)
* ZERO SET (Zero calibration triggered on the transition
from inactive to active on the control input)
* RESET TOTAL (Totaliser reset to zero on the transition
from inactive to active on the control input)
* CLEAR. MSG. (Status warnings cleared on the
transition from inactive to active on the control input)
3.6.2
ACTIV.LEVEL
Set the desired voltage level for the input to be active
* ACTIVE LOW (0 to 2 V)
* ACTIVE.HIGH (4 to 24 V)
3.7.0
SYS.CTRL S
Submenu 3.7.0 System control
3.7.1
FUNCTION S
Function for system control
* OFF (System control inactive)
* FLOW = OFF (Mass flow readings forced to zero,
totaliser frozen)
* FLOW = 0/RST. (Mass flow readings forced to zero,
totaliser frozen while active but reset to zero
as condition becomes inactive. Not available with
Custody Transfer Protection)
* OUTPUTS OFF (Forces all outputs to their OFF states)
3.7.2
REFERENCE
Condition for triggering the above function
* DENSITY (Function is triggered if density goes outside
Max or Min limits as set in Fcts 3.7.3 and 3.7.4)
* TEMPERATUR (Function is triggered if temperature
goes outside Max or Min limits as set in Fct 3.7.3 and
3.7.4) Function not available with Custody Transfer
Protection.
3.7.3
MIN. LIMIT.
Minimum allowable value of temperature or density
selected in Fct. 3.7.2
Units: depend on function but will correspond
to those set in Fct. 3.2.1 and 3.2.5
Function not available with Custody Transfer Protection.
3.7.4
MAX. LIMIT.
Maximum allowable value of temperature or density
selected in Fct. 3.7.2
Units: depend on function but will correspond
to those set in Fct. 3.2.1 and 3.2.5
Function not available with Custody Transfer Protection.