KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 38

background image

38

Fct. No.

Text

Description and settings

3.6.0

CTRL.INP.E

Submenu 3.6.0 Control input

3.6.1

FUNCTION E

Function of the control input

* OFF (control input inactive)

* STANDBY (When active converter switches to

STANDBY)

* ZERO SET (Zero calibration triggered on the transition

from inactive to active on the control input)

* RESET TOTAL (Totaliser reset to zero on the transition

from inactive to active on the control input)

* CLEAR. MSG. (Status warnings cleared on the

transition from inactive to active on the control input)

3.6.2

ACTIV.LEVEL

Set the desired voltage level for the input to be active

* ACTIVE LOW (0 to 2 V)

* ACTIVE.HIGH (4 to 24 V)

3.7.0

SYS.CTRL S

Submenu 3.7.0 System control

3.7.1

FUNCTION S

Function for system control

* OFF (System control inactive)

* FLOW = OFF (Mass flow readings forced to zero,

totaliser frozen)

* FLOW = 0/RST. (Mass flow readings forced to zero,

totaliser frozen while active but reset to zero

as condition becomes inactive. Not available with

Custody Transfer Protection)

* OUTPUTS OFF (Forces all outputs to their OFF states)

3.7.2

REFERENCE

Condition for triggering the above function

* DENSITY (Function is triggered if density goes outside

Max or Min limits as set in Fcts 3.7.3 and 3.7.4)

* TEMPERATUR (Function is triggered if temperature

goes outside Max or Min limits as set in Fct 3.7.3 and

3.7.4) Function not available with Custody Transfer

Protection.

3.7.3

MIN. LIMIT.

Minimum allowable value of temperature or density

selected in Fct. 3.7.2

Units: depend on function but will correspond

to those set in Fct. 3.2.1 and 3.2.5

Function not available with Custody Transfer Protection.

3.7.4

MAX. LIMIT.

Maximum allowable value of temperature or density

selected in Fct. 3.7.2

Units: depend on function but will correspond

to those set in Fct. 3.2.1 and 3.2.5

Function not available with Custody Transfer Protection.