beautypg.com

Chapter 6 status information, 1 status information and filter, 1 status information and filter -1 – Yokogawa µR20000 User Manual

Page 91: App index

background image

6-1

IM 04P01B01-17E

Status

Information

1

2

3

4

5

6

App

Index

6.1 Status Information and Filter

The following figure illustrates the status information and filter on the recorder.

Status information

Filter

Condition register

4 bytes

4

3

2

1

4

3

2

1

• When a status indicated on the following page is entered, the corresponding bit in the

condition register is set to “1.” The logical AND of the condition register and the filter
becomes the status information.

• The IF command can be used to set the filter.
• The IS command is used to output the status information. Status information 1 and 2

are cleared when they are output. Status information 3 and 4 are not cleared when it
is output, and remains at “1” while the event is occurring.

• When multiple connections are up, filters can be specified for the individual

connection. Therefore, the status information can be held for each connection.

Chapter 6

Status Information

This manual is related to the following products: