beautypg.com

App index, Sc sets the chart speed, Sd sets the date and time – Yokogawa µR20000 User Manual

Page 56: Vt sets the dot printing interval (dot model), Sz sets zone recording, Sp sets the partial expanded recording

background image

Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-13

SC

Sets the chart speed.

Syntax

SC p1
p1 Chart speed

Query

SC?

Example Set the chart speed to 25 mm/h.

SC25

Description Select the chart speed from the list of choices

below.
Pen model
5 to 12000 mm/h (82 levels, see section 4.3)
Dot model
1 to 1500 mm/h (1 mm steps)

SD

Sets the date and time.

Syntax

SD p1
p1 Date and time (fixed to the YY/MM/DD_HH:

MM:SS format)
YY

Year (00 to 99, the lower 2
digits)

MM

Month (01 to 12)

DD

Day (01 to 31)

_

Space

HH

Hour (00 to 23)

MM

Minute (00 to 59)

SS

Second (00 to 59)

Query

SD?

Example Set the recorder clock to 13:00:00, December 1,

2004.
SD04/12/01 13:00:00

Description • The format of p1 is fixed to 17 characters. An

error results if a space is inserted in positions
other than the specified positions.

• On the µR10000 with system version 1.02 or

earlier, the parameter p2 can be used.

p2 Standard time or DST (winter, summer)

Winter Standard time
Summer DST

VT

Sets the dot printing interval

(dot model).

Syntax

VT p1
p1 Dot printing interval setting.

AUTO

Automatically adjust the dot

printing interval according to

the chart speed.

FIX

Record at the fastest printing

interval.

Query

VT?

Example Record at the fastest printing interval.

VTFIX

Description When trend recording is performed on 6

channels, the fastest printing interval is 10
seconds.

SZ

Sets zone recording.

Syntax

SZ p1,p2,p3
p1 Channel number
p2 Leftmost position of the zone (µR10000: 0 to

95, µR20000: 0 to 175) [mm]

p3 Rightmost position of the zone (µR10000: 5

to 100, µR20000: 0 to 180) [mm]

Query

SZ[p1]?

Example Display channel 02 in a zone between 30 and 50

mm.
SZ02,30,50

Description • Be sure that p3 is greater than p2 and that the

zone width (p3 – p2) is greater than or equal
to 5 mm.

• On models with the computation function

(/M1 option), computation channels can be
specified.

SP

Sets the partial expanded

recording.

Syntax

SP p1,p2,p3,p4
p1 Channel number
p2 Partial expanded recording ON/OFF (ON,

OFF)

p3 Boundary position (1 to 99) [%]
p4 Boundary value

Query

SP[p1]?

Example Set the 25% position of channel 01 to 1.000 V.

SP01,ON,25,1000

Description • When the input range is set to SKIP (SR

command), p2 cannot be turned ON.

• Set p3 as a percentage where 100 mm is

assumed to be 100%.

• Parameter p4 can be set in the range (leftmost

value of span + 1) to (rightmost value of span
– 1). If scaling is enabled, the range is (leftmost
value of scaling – 1) to (rightmost value of
scaling + 1). Set the parameter using an
integer.

• This setting is possible when partial expanded

recording is enabled (UK command).

• This command cannot be used if the partial

expanded recording range does not exist
(when the span width is set to 1, for example).

• On models with the computation function

(/M1 option), computation channels can be
specified.

• When the computation equation setting (SO

command) is turned OFF, p2 cannot be turned
ON.

4.4 Setting Commands

This manual is related to the following products: