Query
SJ[p1]?
Example Enable timer 1 on computation channel 0B. No
sum scale designation.
SJ0B,1,OFF
Description • This command cannot be specified while
computation is in progress.
• About p3
The data for sum computation are sampled
every scan interval. For data with units such
as /s, /min, /h, and /day as in a flow rate, the
data can be summed over the unit time as
shown below. Parameter p3 is valid only for
sum values.
OFF Σ (measured value)
/s
Σ (measured value) × scan interval
/min Σ (measured value) × scan interval/
60
/h
Σ (measured value) × scan interval/
3600
/day Σ (measured value) × scan interval/
(3600×24)
The unit of the scan interval is seconds.
VD
Sets the data display screen.
On screens without detailed parameters
Syntax
VD p1,p2
p1 Screen number (01 to 15)
p2 Display type
4CH bargraph
4-channel bar graph display (pen
model)
6CH bargraph
6-channel bar graph display (µR10000
dot model)
Channel alarm status
Channel alarm status display
Time/Chart speed
Date/Time/Chart speed display
DI/DO
DI/DO status display
STATUS
Status display
System
System display
Batch name
Batch name (/BT1 option)
Lights out
Screen OFF
SKIP
Skip
Query
VD[p1]?
Example Assign the status display to screen 02.
VD02,STATUS
On screens with the displayed channel switching
interval
Syntax
VD p1,p2,p3
p1 Screen number (01 to 15)
p2 Display type
1CH digital
1-channel digital display
2CH digital
2-channel digital display
4CH digital
4-channel digital display
6CH digital
6-channel digital display (dot model)
12CH digital
12-channel digital display (µR20000
12, 18 and 24 dot model)
1CH digital+1CH bargraph
1-channel digital display + 1 channel
bar graph display
1CH digital+4CH bargraph
1-channel digital display + 4 channel
bar graph display (pen model)
2CH digital+2CH bargraph
2-channel digital display + 2 channel
bar graph display
Tag_1CH digital
1-channel digital display (tag)
Tag_2CH digital
2-channel digital display (tag)
Tag_1CH digital+1CH bargraph
1-channel digital display + 1 channel
bar graph display (tag)
Tag_1CH digital+4CH bargraph
1-channel digital display + 4 channel
bar graph display (tag, pen model)
p3 Displayed channel switching interval
MANUAL Fixed display channels, data
update rate (pen model): 2 s,
data update rate (dot model):
scan interval
AUTO1S Switch the displayed channels
and data every second.
AUTO2S Switch the displayed channels
and data every 2 seconds.
AUTO3S Switch the displayed channels
and data every 3 seconds.
AUTO4S Switch the displayed channels
and data every 4 seconds.
AUTO5S Switch the displayed channels
and data every 5 seconds.
Query
VD[p1]?
4.4 Setting Commands