Status information, User information, Status information -12 user information -12 – Yokogawa µR20000 User Manual
Page 88
5-12
IM 04P01B01-17E
Status Information
• The IS command is used to output the data.
• The operation status of the recorder is output
• For details on the status information, see section 6.2, “The Bit Structure of the Status
Information.”
• Syntax
EACRLF
ddd.ccc.bbb.aaaCRLF
ENCRLF
aaa Status information 1 (000 to 255)
bbb Status information 2 (000 to 255)
ccc Status information 3 (000 to 255)
ddd Status information 4 (000 to 255)
• Example
EA
000.000.032.000
EN
User Information
• The FU command is used to output the data.
• User name, user level, and other information are output.
• Syntax
EACRLF
p_l_uuu···uCRLF
ENCRLF
p Physical layer
E: Ethernet
S: RS-422A/485
l User level
A: Administrator
U: User
uuu···u User name (up to 16 characters)
_ Space
• Example
EA
E A admin
EN