beautypg.com

App index, Ug sets the setting mode menu selection, Uh sets the func key menu selection – Yokogawa µR20000 User Manual

Page 70: Ye exits from basic setting mode, Xe exits from basic setting mode

background image

Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-27

When p1 =

FULL

µR10000: An integer between –30 to 30

(reference value: 1000)

µR20000: An integer between –50 to 50

(reference value: 1800)

Example Adjust the 100% position. Set the record position

adjustment value to 25.
UAFULL,25

Description • Carry out the adjustment in the following order:

P1 = Hysteresis, ZERO, and then FULL.

• Check the recorded result and correct the

record position adjustment value.

• To end the adjustment, execute the UY0

command. If you do not end the adjustment,
you cannot change the execution mode.

• The unit for p2 is dot (0.1 mm).

UG

Sets the Setting mode menu

selection.

Syntax

UG p1,p2,p3,p4,p5,p6,p7,p8,p9,p10

p1 Range (ON, OFF)
p2 Bias (ON, OFF)
p3 Alarm (ON, OFF)
p4 Units (ON, OFF)
p5 Chart speed (ON, OFF)
p6 Other (ON, OFF)
p7 Calibration correction (ON, OFF) (/CC1

option)

p8 Computation (ON, OFF) (/M1 option)
p9 Batch name (ON, OFF) (/BT1 option)
p10 Batch details (ON, OFF) (/BT1 option)

Query

UG?

Example Display all menus.

UGON,ON,ON,ON,ON,ON,ON,ON,ON,ON

Description • Parameter p7 can be specified on models with

the calibration correction (/CC1 option).

• Parameter p8 can be specified on models with

the computation function (/M1 option).

• Parameters p9 and p10 can be specified on

models with the header printout (/BT1 option).

Note

Parameters p7 and higher are skipped when no options are

installed.

UH

Sets the FUNC key menu

selection.

Syntax

UH p1,p2,p3,p4,p5
p1 Manual printout (ON, OFF)
p2 Setup printout (ON, OFF)
p3 Message printout (ON, OFF)
p4 Buffer clear (ON, OFF)
p5 Periodic printout (ON, OFF)

Query

UH?

Example Display all menus.

UHON,ON,ON,ON,ON

UE

Selects enable/disable for

Start printout/End printout and

message format (/BT1 option).

When using Start printout/End printout

Syntax

UE p1,p2,p3,p4
p1 Use of Start printout/End printout (USE)
p2 Number of digits for lot number (4, 6, NOT)
p3 Use/Not use Start printout 2/End printout 2

(USE, NOT)

p4 Use/Not use message format (USE, NOT)

Query

UE?

Example Perform Start printout/End printout, set 6 digits

for lot number, use Start printout 2/End printout 2
and message format.
UEUSE,6,USE,USE

Description • Set the lot number using the VH command.

• Set the message format using the VM

command.

When not using Start printout/End printout

Syntax

UE p1,p2
p1 Start printout/End printout not used (NOT)
p2 Use/Not use message format (USE, NOT)

Query

UE?

Example Use message format but do not use Start

printout/End printout.
UENOT,USE

Description Set the message format using the VM command.

YE

Exits from Basic Setting mode.

When the YE command is executed, communication is
dropped.

Syntax

YE p1
p1 Apply/Discard the settings

STORE

Apply the settings

ABORT

Discard the settings

Example Apply the basic settings.

YESTORE

Description Applies the settings changed with the basic

setting commands.

XE

Exits from Basic Setting mode.

Communication is not dropped when the XE command is
executed.
Syntax

XE p1
p1 Apply/Discard the settings

STORE Apply the settings
ABORT Discard the settings

Example Apply the basic settings.

XESTORE

Description The settings specified by the following

commands and saved using the XE command
take effect after the recorder is power cycled.

YS, YB, YA, YN, YD, YQ, and YK

4.5 Basic Setting Commands

This manual is related to the following products: