Rs-422a/485 dedicated commands and responses, Rs-422a/485 dedicated commands and responses -5 – Yokogawa µR20000 User Manual
Page 81
5-5
IM 04P01B01-17E
Responses
1
2
3
4
5
6
App
Index
RS-422A/485 Dedicated Commands and Responses
The following table shows dedicated commands for the RS-422A/485 interface and their
responses.
Command Syntax
Meaning
Response
ESC
Oxx CRLF
Open the instrument.
• Response from the instrument with
the specified address
ESC
Oxx CRLF
• Response when the instrument with the
specified address does not exist*
ESC
Cxx CRLF
Close the instrument.
• Response from the instrument with the
specified address
ESC
Cxx CRLF
• Response when the instrument with the
specified address does not exist*
* The causes that the condition become “The instrument with the specified address does
not exist” is such as a command error, the address not matching that of the instrument, the
instrument is not turned ON, and the instrument not being connected via the serial interface.
• The “
xx” in the table indicates the instrument’s address. Specify the address that is
assigned to the instrument from 01 to 32.
• Only one instrument can be opened at any one time.
• When an instrument is opened with the ESC O command, all commands on the
instrument become active.
• When an instrument is opened with the ESC O command, any other instrument
that is open is automatically closed.
• Use CR+LF for the terminator. LF cannot be used.
Note
The ASCII code of ESC is 1BH.
5.1 Response Syntax