beautypg.com

Measured/computed data, Measured/computed data -8 – Yokogawa µR20000 User Manual

Page 84

background image

5-8

IM 04P01B01-17E

Measured/computed Data

• The FD (FD0) or FY (FYInst, FYTlog1, or FYTlog2) command is used to output the

data.

• Syntax

The measured/computed data are output in the following syntax along with the date

and time information for each channel
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mi:ss.mmmt_S

1

S

2

S

3

S

4

S

5

S

6

CRLF

s_kcca

1

a

2

a

3

a

4

uuuuuuf

1

dddddEf

2

ppCRLF

····························
ENCRLF

yy Year (00 to 99)
mo Month (01 to 12)
dd Day (01 to 31)
hh Hour (00 to 23)
mi Minute (00 to 59)
ss Second (00 to 59)
mmm Millisecond (000 to 999. A period is placed between seconds and

milliseconds.)

t ‘S’(=summer) or ‘ ’ space(=winter)
S

1

S

2

S

3

S

4

S

5

S

6

Data status

Takes on the values below for the FYTlog1 and FYTlog2
commands. For all other cases, they are all spaces.

S

1

Time change during TLOG computation: T (occurred) or space (not

occurred)

S

2

Power OFF and ON during TLOG computation: P (occurred) or

space (not occurred)

S

3

Data reset during the TLOG computation: R (occurred) or space

(not occurred)

S

4

S

5

S

6

All spaces

s Channel data status (N, D, S, O, B, E)

N: Normal
D: Differential input
S: Skip
O: Over
B: Burnout
E: Error

k Channel type

0: Measurement channel
A: Computation channel

cc Channel number

µR10000: 01 to 06, 0A to 0P
µR20000: 01 to 24, 0A to 1P

a

1

a

2

a

3

a

4

a

1

Alarm status (level 1)

a

2

Alarm status (level 2)

a

3

Alarm status (level 3)

a

4

Alarm status (level 4)

5.2 Output Format of ASCII Data