beautypg.com

Terminal arrangement and signal names, Connection procedure, Cable – Yokogawa µR20000 User Manual

Page 30: Connecting the cable, Four-wire system two-wire system

background image

3-2

IM 04P01B01-17E

3.2 Terminal Arrangement and Signal Names and

the Connection Procedure of the RS-422A/485

Communication Interface

Terminal Arrangement and Signal Names

L

N

ALARM

CHART

REMOTE
1 2 3

RS-422A

SD SD SG

A B G

RD RD FG

A B G

4 5 C

NO C NC

NO C NC

NO C NC

NO C NC

NO C NC

NO C NC

RS-422A

SD SD SG

A B

RD RD FG

A B

SDA SDB SG

RDA RDB FG

Rear panel

Terminal Name

Description

FG (Frame Ground)

Case ground of the recorder.

SG (Signal Ground)

Signal ground.

SDB (Send Data B)

Send data B (+).

SDA (Send Data A)

Send data A (–).

RDB (Received Data B)

Receive data B (+).

RDA (Received Data A)

Receive data A (–).

Connection Procedure

• Cable

Use the cable that meets the conditions below.

Item

Conditions

Cable

Shielded twisted pair cable

3 pairs AWG24-14 (Four-wire system),

2 pairs 24 AWG or more (Two-wire system)

Characteristic impedance

100 Ω

Capacitance

50 pF/m

Cable length

Up to 1.2 km*

* The transmission distance of the RS-422A/485 interface is not the straight-line distance, but

rather the total length of the (shielded twisted-pair) cable.

• Connecting the Cable

As shown in the following figure, attach a crimp-on lug with isolation sleeves for 4 mm

screws to the end of the cable. Keep the exposed section from the end of the shield
within 5 cm.

SDB

SDA

SG

RDB

RDA

FG

Electric

potential of

the shield

Shield

Four-wire system

Two-wire system

SDB

SDA

SG

RDB

RDA

FG

Electric

potential of

the shield

Shield

This manual is related to the following products: