Cm sets the communication input data (/m1 option), Vc sets the batch comment (/bt1 option) – Yokogawa µR20000 User Manual
Page 61
4-18
IM 04P01B01-17E
Query
VD[p1]?
Example Assign the display in which different screens can
be assigned to the top and bottom sections to
screen 09. Set the top section to status display
and the bottom section to 1-channel digital
display. Switch the displayed channel every 3
seconds (parameters p5 to p7, p10, and p12
explained above are omitted).
VD09,Multiple display,STATUS,1CH
digital,AUTO3S
CM
Sets the communication input
data (/M1 option).
Syntax
CM p1,p2
p1 Communication input data number (see
section 4.3)
p2 Communication input data
The selectable range is –9.9999E+29 to –
1.0000E–30, 0, 1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
CM?
Example Set 1.0000E–10 to communication input data
number C01.
CMC01,1.0000E−10
FR
Sets the interval for acquiring
data to the FIFO buffer.
Syntax
FR p1
p1 Pen model: 125ms, 250ms, 500ms, 1s, 2s,
2.5s, 5s, 10s
Dot model: 1s, 2s, 2.5s, 5s, 10s
Query
FR?
Example Set the acquiring interval to the FIFO buffer to 1 s.
FR1s
Description • Set the acquiring interval to an integer multiple
of the scan interval that is greater than the
scan interval.
• The scan interval on the pen model is 125 ms.
• The scan interval on the dot model varies
depends on the integration time of the A/D
converter as follows:
When the integration time is 16.7 ms or 20 ms;
6 channels: 1 s
12, 18, or 24 channels: 2.5 s
When the integration time is 100 ms;
6 channels: 2.5 s
12 channels: 5 s
18 and 24 channels: 10 s
VH
Sets the batch number and lot
number (/BT1 option).
When setting the batch number
Syntax
VH p1,p2
p1 BATCH
p2 Batch number (up to 26 characters)
Query
VH[p1]?
Example Set character string Product as a batch number.
VHBATCH,Product
Description Valid with the header printout (/BT1 option) when
set to use Start printout/End printout (see the UE
command).
When setting lot number for the 4 digits
Syntax
VH p1,p2
p1 LOT
p2 0000-9999
Query
VH[p1]?
Example Set 0001 as the lot number.
VHLOT,0001
Description Valid with the header printout (/BT1 option)
when lot number is set to 4 digits (see the UE
command).
When setting lot number for the 6 digits
Syntax
VH p1,p2
p1 LOT
p2 000000-999999
Query
VH[p1]?
Example Set 000001 as the lot number.
VHLOT,000001
Description Valid with the header printout (/BT1 option)
when lot number is set to 6 digits (see the UE
command).
VC
Sets the batch comment
(/BT1 option).
Syntax
VC p1,p2,p3
p1 Mode selection
START Start printout
END
End printout
START2 Start printout 2
END2
End printout 2
p2 line number (up to 5)
p3 Batch comment
(µR10000: up to 32 chars., µR20000: up to
64 chars.)
Query
VC[p1[,p2]]?
Example Set the string Batch start to the first line of the
Start printout batch comment.
VCSTART,1,Batch start
Description • Valid with the header printout (/BT1 option)
when set to use Start printout/End printout (see
the UE command).
• Start printout 2 and End printout 2 valid when
set to use Start printout 2 and End printout 2
(see the UE command).
4.4 Setting Commands