• Parameter p5 can be specified on models

with the calibration correction (/CC1 option).

Use the UQ and VL commands to set the

correction values.

UT

Selects the time printout format.

Syntax

UT p1,p2,p3,p4
p1 Time printout format of the alarm printout

The available selections are the same as
those of p2. None is not available.

p2 Time printout format of the message printout

HH:MM Hour:Minute
HH:MM:SS Hour:Minute:Second
M/D H:M Month Day Hour:Minute
M/D H:M:S Month Day Hour:Minute:

Second

YMD H:M:S Month Day Year Hour:Minute:

Second

None

No time printout

p3 Time printout format of the record start time

printout

The available selections are the same as
those of p2. None is not available.

p4 Time printout format of the new chart speed

printout

The available selections are the same as
those of p2. None is not available.

Query

UT?

Example Set the all time printouts to Hour:Minute format.

UTHH:MM,HH:MM,HH:MM,HH:MM

XR

Sets the remote control input

(/R1 option).

Syntax

XR p1,p2
p1 Remote control input terminal number (1 to

5)

p2 Action

NONE

No action

Record On/Off

Start/Stop recording.

Chart speed

Change the chart
speed.

Time adjust

Reset the internal
clock to the nearest
hour.

Math start/stop Start/Stop

computation.

Math reset

Reset computation.

Manual print

Execute manual
print.

Alarm Ack

Execute alarm
acknowledge.

Message1

Print message 1

Message2

Print message 2

Message3

Print message 3

Message4

Print message 4

Message5

Print message 5

Priority R_RCD

Priority to remote
recording

BatchCMT switch

Batch comment
switching

Query

XR[p1]?

Example Assign the printout of message 2 to the remote

control input of terminal number 1.
XR1,Message2

Description • Math start/stop and Math reset can be

specified on models with the computation
function (/M1 option).

• “

Priority R_RCD” and “BatchCMT

switch” can be set on products with header
printout (/BT1 option).

YS

Sets the RS-422A/485 interface

(/C3 option).

Syntax

YS p1,p2,p3,p4,p5
p1 Address (1 to 32)
p2 Baud rate (1200, 2400, 4800, 9600, 19200,

38400)

p3 Data length (7, 8)
p4 Parity check (ODD, EVEN, NONE)
p5 Protocol (NORMAL, MODBUS)

Query

YS?

Example Set the address to 2, the baud rate to 9600, the

data length to 8, the parity check to ODD, and
the NORMAL protocol.
YS2,9600,8,ODD,NORMAL

Description • The settings specified by this command and

saved using the XE command take effect after
the recorder is power cycled.

• If p5 is set to MODBUS, Modbus slave is

selected.

XQ

Sets the TLOG timer

(/M1 option).

When not using the timer

Syntax

XQ p1,p2
p1 Timer number (1 or 2)
p2 Timer type (OFF)

Query

XQ[p1]?

Example Disable the number 1 timer.

XQ1,OFF

When using an absolute timer

Syntax

XQ p1,p2,p3,p4,p5,p6
p1 Timer number (1 or 2)
p2 Timer type (ABSOLUTE)
p3 Interval (10min, 12min, 15min, 20min,

30min, 1h, 2h, 3h, 4h, 6h, 8h, 12h, 24h)

p4 Reference time (fixed to HH format)

HH

Hour (00 to 23)

p5 Reset/not reset the data when the timer

expires (ON/OFF)