beautypg.com

Report data generated by the periodic printout, Report data generated by the periodic printout -10 – Yokogawa µR20000 User Manual

Page 86

background image

5-10

IM 04P01B01-17E

Report Data Generated by the Periodic Printout

• The FY command (FYREPORT) is used to output the data.
• Report data generated by the periodic printout is output.

Syntax

EACRLF
YY/MO/DD_HH:MI:SS.MMMTCRLF
yy/mo/dd_hh:mi:ss.mmmt_S

1

S

2

S

3

S

4

S

5

S

6

CRLF

s

1

s

2

s

3

s

4

s

5

s

6

kccuuuuuuf

1

dddddEf

2

pp_f

1

dddddEf

2

pp_f

1

dddddEf

2

pp_f

1

ddddd

Ef

2

pp_f

1

ddddddddEf

2

ppCRLF

··················
ENCRLF

YY/MO/DD HH:MI:SS.MMMT

Report start time information

yy/mo/dd hh:mi:ss.mmmt

Report end time information

YY, yy Year (00 to 99)
MO, mo Month (01 to 12)
DD, dd Day (01 to 31)
HH, hh Hour (00 to 23)
MI, mi Minute (00 to 59)
SS, ss Second (00 to 59)
MMM, mmm Millisecond (000 to 999)
T, t ‘S’(=summer) or ‘ ’ space(=winter)

S

1

S

2

S

3

S

4

S

5

S

6

Data status

S

1

Time change during reporting: T (occurred) or space (not

occurred)

S

2

Power OFF and ON during reporting: P (occurred) or space (not

occurred)

S

3

Data clear during reporting: R (occurred) or space (not occurred)

S

4

S

5

S

6

All spaces

s

1

s

2

s

3

s

4

s

5

s

6

Channel data status

s

1

Channel mode at the end of reporting: S (Skipped) or space

(Mode other than skipped)

s

2

Range change during reporting: C (occurred) or space (not

occurred)

s

3

Error data occurrence during reporting: E (yes) or space (no)

s

4

±over data occurrence during reporting: O (yes) or space (no)

s

5

s

6

All spaces

cc Channel number

µR10000: 01 to 06, 0A to 0P
µR20000: 01 to 24, 0A to 1P

k Channel type (0, A)

0: Measurement channel
A: Computation channel

uuuuuu Unit information (6 characters)

5.2 Output Format of ASCII Data