App index – Yokogawa µR20000 User Manual
Page 39
3-11
IM 04P01B01-17E
Using
the
RS-422A/485
Communication
Interface
(/C3
Option)
1
2
3
4
5
6
App
Index
Input Register Data
33001
Alarm status of the computed data of CH0A
33002
Alarm status of the computed data of CH0B
33003
Alarm status of the computed data of CH0C
33004
Alarm status of the computed data of CH0D
33005
Alarm status of the computed data of CH0E
33006
Alarm status of the computed data of CH0F
33007
Alarm status of the computed data of CH0G
33008
Alarm status of the computed data of CH0J
33009
Alarm status of the computed data of CH0K
33010
Alarm status of the computed data of CH0M
33011
Alarm status of the computed data of CH0N
33012
Alarm status of the computed data of CH0P
33013
Alarm status of the computed data of CH1A
33014
Alarm status of the computed data of CH1B
33015
Alarm status of the computed data of CH1C
33016
Alarm status of the computed data of CH1D
33017
Alarm status of the computed data of CH1E
33018
Alarm status of the computed data of CH1F
33019
Alarm status of the computed data of CH1G
33020
Alarm status of the computed data of CH1J
33021
Alarm status of the computed data of CH1K
33022
Alarm status of the computed data of CH1M
33023
Alarm status of the computed data of CH1N
33024
Alarm status of the computed data of CH1P
• Registers corresponding to models with the /M1 computation function option.
• The values are the same as those of the alarm status of the measured data.
• Valid channels vary depending on the device. An error response (code 2) occurs when an
invalid channel is read.
Input Register Data
36001
List of alarms of the measured data of CH01 to CH04
36002
List of alarms of the measured data of CH05 to CH08
36003
List of alarms of the measured data of CH09 to CH12
36004
List of alarms of the measured data of CH13 to CH16
36005
List of alarms of the measured data of CH17 to CH20
36006
List of alarms of the measured data of CH21 to CH24
36007 to 36020 Always 0
36021
List of alarms of the computed data of CH0A to CH0D
36022
List of alarms of the computed data of CH0E to CH0J
36023
List of alarms of the computed data of CH0K to CH0P
36024
List of alarms of the computed data of CH1A to CH1D
36025
List of alarms of the computed data of CH1E to CH1J
36026
List of alarms of the computed data of CH1K to CH1P
1 register (1 word)
Level 1
Level 2
Level 4
Level 3
The bit is set to 1
when an alarm occurs.
Example of Register 36001
CH4
CH3
CH2
CH1
• Registers 36001 to 36026 can be read with a single command.
• Bits corresponding to invalid channels are fixed to “0.”
3.4 Modbus Slave Specifications