Commands

1

2

3

4

5

6

App

Index

IM 04P01B01-17E

4-31

Newest value output (GETNEW)

Output the specified number of blocks (p4)

of FIFO data back starting from the recent
acquire position (block).

• Parameters p2, p3, and p4 are valid when p1

is set to GET or GETNEW.

• If p4 is omitted, all the data of all blocks

acquired in the FIFO buffer are output.

• Be sure to set p2 and p3 so that p3 is greater

than or equal to p2.

• For the output flow of FIFO data, see appendix

4.

IS

Outputs status information.

Syntax

IS p1
p1 Output status information (0)

Example Output status information.

IS0

Description The output status can be masked using the

status filter (IF command).

FU

Outputs user information.

Syntax

FU p1
p1 Output user information (0)

Example Output user information.

FU0

Description Outputs the information of the user currently

connected to the recorder.

4.8 RS-422A/485 Dedicated

Commands

ESC O Opens the instrument.

The ASCII code of ESC is 1BH.

Syntax

ESC O p1
p1 Instrument address (01 to 32)

Example Open the instrument at address 01.

ESC O01

Description • Specifies the address of the instrument with

which to communicate.

• Only one instrument can be opened at any

given time.

• When an instrument is opened with the ESC

O command, any other instrument that is
currently open is automatically closed.

• When this command is received correctly, the

recorder transmits the data “ESC 0 xx” (xx:
Instrument’s address). For the description of
the response, see page 5-5.

• Use CR+LF for the terminator. LF cannot be

used.

ESC C Closes the instrument.

The ASCII code of ESC is 1BH.

Syntax

ESC C p1
p1 Instrument address (01 to 32)

Example Close the instrument at address 01.

ESC C01

Description • Clears the current connection with the

instrument.

• When this command is received correctly, the

recorder transmits the data “ESC C xx” (xx:
Instrument’s address). For the description of
the response, see page 5-5.

• Use CR+LF for the terminator. LF cannot be

used.