App index, Tw sets the secondary trend interval, Tm sets manual sampling – Yokogawa Button Operated MV2000 User Manual

Page 87: Te sets sampling conditions of the event data, Sz sets a zone

background image

4-13

Commands

IM MV1000-17E

1

2

3

4

5

6

7

App

Index

SW

Sets the trend/strage interval and

auto-save interval

Syntax

SW p1,p2,p3,p4
p1 1
p2 Waveform type (T-Y)
p3 Trend/strage interval (5S, 10S, 15S, 30S,

1MIN, 2MIN, 5MIN, 10MIN, 15MIN, 20MIN,
30MIN, 1H, 2H, 4H, 10H)

p4 Auto-save interval (10MIN, 20MIN, 30MIN,

1H, 2H, 3H, 4H, 6H, 8H, 12H, 1DAY, 2DAY,
3DAY, 5DAY, 7DAY, 10DAY, 14DAY, 31DAY)

Query

SW?

Description • You cannot execute this command while the

MV is measuring.

• The selectable auto-save interval (p4) varies

depending on the trend/strage interval (p3)
setting. For details, see the MV1000/MV2000
User’s Manual
.

• You can specify the trend/strage intervals (p3)

5S and 10S only on high-speed input models
(MV1004, MV1008, and MV2008). You can
specify 15S on medium-speed input models
set to fast sampling mode and high-speed
input models.

• Set the trend/strage interval (p3) to an interval

slower than the scan interval.

• The p4 setting is valid when the saving

method to the external storage medium is set
to auto with the XM command (p1 in the XM
command set to AUTO).

• Set the trend interval with the NR command

after setting the trend/strage interval and auto
save interval with the SW command.

• The trend/strage interval (p3) can only be set

to an interval slower than the scan interval.
(The scan interval is set using p3 in the XV
command.)

• The selectable range of auto save interval (p4)

varies depending on the trend/strage interval
(p3) setting and the number of channels that
is set with the TJ command.

TW

Sets the secondary trend interval

Syntax

TW p1
p1 Trend interval (5S, 10S, 15S, 30S, 1MIN,

2MIN, 5MIN, 10MIN, 15MIN, 20MIN, 30MIN,
1H, 2H, 4H, 10H)

Query

TW?

Example Set the trend interval to 2 minutes.

TW2MIN

Description • Set the trend interval (p1) to an interval slower

than the scan interval.

• You can specify the trend intervals (p1) 5S

and 10S only on high-speed input models
(MV1004, MV1008, and MV2008). You can
specify 15S on medium-speed input models
set to fast sampling mode and high-speed
input models.

TM

Sets manual sampling.

Syntax

TM p1,p2,p3
p1 Manual sampling channel number
p2 Enable/Disable (ON, OFF)
p3 Measurement, computation, or external

input channel number

Query

TM[ p1]?

Example Assign measurement channel 002 to manual

sampling number 001.
TM001,ON,002

Description • You can execute this command on models

with the /MC1 external input channel option.

• You can assign a computation channel on

models with the /M1 math option.

TE

Sets sampling conditions of the

event data

Syntax

TE p1,p2,p3,p4,p5,p6
p1 1
p2 Sample rate (25MS, 125MS, 250MS,

500MS, 1S, 2S, 5S, 10S, 30S, 1MIN, 2MIN,
5MIN, 10MIN)

p3 Sample mode

FREE

Starts data acquisition at memory
start and stops data acquisition at
memory stop.

SINGLETRIGGER

Acquires data for a specified time
once after a trigger occurs and
stops.

REPEATTRIGGER

Acquires data for a specified time
after a trigger occurs and waits for
the next trigger.

p4 Sample time (10MIN, 20MIN, 30MIN, 1H,

2H, 3H, 4H, 6H, 8H, 12H, 1DAY, 2DAY,
3DAY, 5DAY, 7DAY, 10DAY, 14DAY, 31DAY)

p5 Pre-trigger length (0, 5, 25, 50, 75, 95, 100)

percent

p6 Key trigger source ON/OFF

Parameters p5 to p6 are valid when p3 is set to
SINGLETRIGGER or REPEATTRIGGER.

Query

TE[ p1]?

Example Acquire data at a sampling rate of 125-ms over

10 minutes using a single trigger.
TE1,125MS,SINGLETRIGGER,10MIN

Description You cannot specify a sampling rate that is faster

than the scan interval.

SZ

Sets a zone

Syntax

SZ p1,p2,p3
p1 Measurement, computation, or external

input channel number

p2 Bottom edge of zone (0 to 95) [%]
p3 Top edge of zone (5 to 100) [%]

4.4 Setting Commands (Setting)