<6 OPERATION>

6-38

IM 01R01A11-01EN

6.3.4

Insulation Resistance

WARNING

Follow User’s Manual IM 01E20D01-01E for correct operational and safe handling.
Where the insulation resistance check is not possible, select “N/A”

1) Check the insulation resistance in the terminal box in accordance with the tables below. Fol-

low User’s Manual IM 01E20D01-01E.

(A) Coil

Table 6.3.13 Coil Insulation Resistance Check

Test Terminals

Test Voltage

Between terminals EX1 and C

500 V DC (Use an insulation tester or equivalent.)

(B) Signal (Electrode)

Table 6.3.14 Electrode Insulation Resistance Check

Test Terminals

Test Voltage

- Between terminals A and C

- Between terminals B and C

500 V DC (Use an insulation tester or equivalent.)

2) Enter the measured value manually.

Where the insulation resistance check is not possible, select “N/A”.

Where the measured value is over-range, select “Open Loop (O.L.)”.

3) “Result” is displayed automatically.

F060319-A.ai

(A) Example 1

F060319-B.ai

(B) Example 2

Figure 6.3.19 “Insulation Resistance” Result