<6 OPERATION>

6-35

IM 01R01A11-01EN

6.3.3

Converter

WARNING

Follow User’s Manual FY2-XJAM012-40-2E for correct operation and safe handling.

This item checks the accuracy of the AXF converter with Magnetic Flowmeter Calibrator AM012.

1

2

3

5

4

F060315.ai

Figure 6.3.15 “Converter” Screen

Table 6.3.9 Converter

Item

No.

Item Name

Description

1

Connection diagram

Shows the connection diagram on another screen. The diagram is (A) or

(B) in

Figure 6.3.16 depending on the case.

2

Span

Shows the Span of the converter.

This field can be edited.

3

Read Span

Reads the Span of the converter. Press “Read Span”.

(Only when Verification Tool is online to AXF)

4

Measured (m/s)

Shows the measured value.

This field can be edited.

5

Read

Reads the measured value. Press “Read”.

(Only when Verification Tool is online to AXF)

SB

B

A

SA

C

AL–

AL+

COM

SO2+

SO1+

I+

I–

CUR OUT

STATUS OUT

ALARM OUT

SIGNAL

COM

SI2+

SI1+

EX2

EX1

L/+

N/–

POWER SUPPLY

EXCITER

P–

P+

PULSE OUT

STATUS IN

(Shield)

(White)

(Black)

(Shield) (White)

(Black)

AXFA11

AM012

F060316-A.ai

(Shield)

(White)

(Black)

(Shield)

(White)

(Black)

N /

L /

+

SA

SB

+

+

+

AM012

Junction Box Terminal

F060316-B.ai

(A) AXFA11

(B) AXFA14 or AXF Integral Flowmeter

Figure 6.3.16 “Converter” Connection Diagram