beautypg.com

2 save, Save -4, A) when “no flow + flow” is not complete – Yokogawa FieldMate User Manual

Page 36

background image

<6 OPERATION>

6-4

IM 01R01A11-01EN

6.1.2

Save

By clicking the “Save” icon, the verifi cation data is saved to the database.
After clicking “Save”, the screen changes as follows:

• When “No Flow + Flow (*1)” is not complete, see (A).
• When “No Flow + Flow” is complete, see (B).

*1: “No Flow + Flow”;

“Circuit” check and “Device Status” check for both “No Flow” and “Flow” conditions. Refer to 6.2.1.

F060105.ai

Figure 6.1.5 “Save” Icon

(A) When “No Flow + Flow” is not complete:

1) After clicking the “Save” icon,

Figure 6.1.6 is displayed.

1

2

3

4

5

F060106.ai

Figure 6.1.6 “Saving Data When “No Flow + Flow” Uncompleted” Dialog

Table 6.1.3 “Saving Data When “No Flow + Flow” Uncompleted” Dialog

Item

No.

Item Name

Description

1

Reason

Shows the reason or comment. It can be used as a marker when loading

verifi cation data is selected. Refer to

Figure 6.1.3.

Enter preferred text.

(Optional fi eld)

2

Saved as Installation Data. Saves the data as “Installation Data”. See 7.1.

This fi eld is disabled until “No Flow + Flow” is complete.

3

Saved as Locked Data.

Saves the data as “Locked Data”. See 7.2.

To save the data as read-only, check the box.

(Always valid)

4

OK

Saves the data.

5

Cancel

Cancels saving the data and returns to “Main Screen”. See 6.1.