6 operation, A) axfa11, Figure 6.3.10 “pulse output” result – Yokogawa FieldMate User Manual
Page 62
<6 OPERATION>
6-30
IM 01R01A11-01EN
NOTE
■ Measure for more than 60 seconds.
■ If using a YOKOGAWA CA150, set the mode to “Contact Input” and the range to “CPM”.
4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or
equivalent to simplify the measurement of pulses. If such a device is unavailable, an appro-
priate load (30V DC, 0.2A max) is required and use the confi guration shown in
Figure 6.3.9.
SB
B
A
SA
C
AL–
AL+
COM
SO2+
SO1+
I+
I–
CUR OUT
STATUS OUT
ALARM OUT
SIGNAL
COM
SI2+
SI1+
EX2
EX1
L/+
N/–
POWER SUPPLY
EXCITER
P–
P+
PULSE OUT
STATUS IN
Electronic Counter
30VDC , 0.2A. max
External
power supply
Load
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
F060309-A.ai
(A) AXFA11
Electronic Counter
30VDC , 0.2A. max
External
power supply
Load
Dashed box : Yokogawa CA150 Handy Calibrator or equivalent
F060309-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.9 Pulse Output Diagram (without optional code EM)
5) Enter the measured value manually.
6) “Result” is displayed automatically.
F060310.ai
Figure 6.3.10 “Pulse Output” Result