2 verification item, Verification item -3 – Yokogawa FieldMate User Manual
Page 11
<2 OUTILINE>
2-3
IM 01R01A11-01EN
2.2
Verification Item
This Verification Tool has 2 modes, which can be run separately.
1. Standard Verification (Standard VF)
• This mode verifies the condition of the AXF without difficulty by checking its internal param-
eters. No external instruments are required.
• Some items are run only when FieldMate is in “online (*1)” mode.
2. Enhanced Verification (Enhanced VF) [Optional]
• This mode provides a more reliable verification by performing a detailed overview equivalent
to SERVICE.
• This mode is optional. If only this mode is completed, “Overall Status” is left blank.
• This mode verifies AXF by using external instruments, such as Magnetic Flowmeter Calibra-
tor AM012, Handy Calibrator CA150, and so on.
• Even when FieldMate is in “offline (*2)” mode, this mode can be run; in this case, some set-
tings must be made manually.
There are only 2 possible combinations:
1. Standard VF
2. Standard VF + Enhanced VF
Table 2.2.1 Verification Items
Mode
Verification Item
Description
Standard VF
Circuit (*3)
Magnetic Circuit
Checks the Magnetic Circuit.
Excitation Circuit
Checks the Excitation Circuit.
Calculation Circuit
Checks the Calculation Circuit.
Device Status (*3)
Alarm Check
Checks occurring Alarms.
Alarm History
Checks historical Alarms.
Physical Appearance
Flowtube
Checks the Flowtube appearance.
Converter
Checks the Converter appearance.
Enhanced VF
Current Output (*4)
Checks the Current Output.
Pulse Output (*4)
Checks the Pulse Output.
Converter (*4)
Checks the Converter Accuracy.
Insulation Resistance
Coil
Checks the Excitation Coil.
(Remote Flowtube Only)
Signal (Electrode)
Checks the Insulation Resistance.
(Remote Flowtube Only)
*1:
Online: The status in which the software communicates with a device.
*2:
Offline: The status in which the software does not communicate with a device.
The communication status between FieldMate and AXF can be confirmed on the FieldMate window.
*3:
These items are checked only when Verification Tool is in “online” mode.
*4:
When Verification Tool is in “offline” mode, manual setting is necessary.