beautypg.com

2 verification item, Verification item -3 – Yokogawa FieldMate User Manual

Page 11

background image

<2 OUTILINE>

2-3

IM 01R01A11-01EN

2.2

Verification Item

This Verification Tool has 2 modes, which can be run separately.
1. Standard Verification (Standard VF)

• This mode verifies the condition of the AXF without difficulty by checking its internal param-

eters. No external instruments are required.

• Some items are run only when FieldMate is in “online (*1)” mode.

2. Enhanced Verification (Enhanced VF) [Optional]

• This mode provides a more reliable verification by performing a detailed overview equivalent

to SERVICE.

• This mode is optional. If only this mode is completed, “Overall Status” is left blank.
• This mode verifies AXF by using external instruments, such as Magnetic Flowmeter Calibra-

tor AM012, Handy Calibrator CA150, and so on.

• Even when FieldMate is in “offline (*2)” mode, this mode can be run; in this case, some set-

tings must be made manually.

There are only 2 possible combinations:

1. Standard VF
2. Standard VF + Enhanced VF

Table 2.2.1 Verification Items

Mode

Verification Item

Description

Standard VF

Circuit (*3)

Magnetic Circuit

Checks the Magnetic Circuit.

Excitation Circuit

Checks the Excitation Circuit.

Calculation Circuit

Checks the Calculation Circuit.

Device Status (*3)

Alarm Check

Checks occurring Alarms.

Alarm History

Checks historical Alarms.

Physical Appearance

Flowtube

Checks the Flowtube appearance.

Converter

Checks the Converter appearance.

Enhanced VF

Current Output (*4)

Checks the Current Output.

Pulse Output (*4)

Checks the Pulse Output.

Converter (*4)

Checks the Converter Accuracy.

Insulation Resistance

Coil

Checks the Excitation Coil.

(Remote Flowtube Only)

Signal (Electrode)

Checks the Insulation Resistance.

(Remote Flowtube Only)

*1:

Online: The status in which the software communicates with a device.

*2:

Offline: The status in which the software does not communicate with a device.

The communication status between FieldMate and AXF can be confirmed on the FieldMate window.

*3:

These items are checked only when Verification Tool is in “online” mode.

*4:

When Verification Tool is in “offline” mode, manual setting is necessary.