beautypg.com

6 operation, A) axfa11, Figure 6.3.14 “active pulse output” result – Yokogawa FieldMate User Manual

Page 65

background image

<6 OPERATION>

6-33

IM 01R01A11-01EN

NOTE

• Measure for more than 60 seconds.
• If using a YOKOGAWA CA150, set the mode to “Frequency (FREQ) and Pulse” and the

range to “CPM”.

• An appropriate load (24 VDC ± 20%, 150 mA max) is required.

4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or

equivalent to simplify the measurement of pulses. If such a device is unavailable, use the

confi guration shown in

Figure 6.3.13.

SB

B

A

SA

C

AL–

AL+

COM

SO2+

SO1+

I+

I–

CUR OUT

STATUS OUT

ALARM OUT

SIGNAL

COM

SI2+

SI1+

EX2

EX1

L/+

N/–

POWER SUPPLY

EXCITER

P–

P+

PULSE OUT

STATUS IN

Electronic Counter

Output voltage: 24 V DC ±20%

Current: 150 mA max

Load

Dashed box : Yokogawa CA150 Handy Calibrator or similar

F060313-A.ai

(A) AXFA11

Electronic Counter

Output voltage: 24 V DC ±20%

Current: 150 mA max

Load

Dashed box : Yokogawa CA150 Handy Calibrator or similar

F060313-B.ai

(B) AXFA14 or AXF Integral Flowmeter

Figure 6.3.13

Active Pulse Output Diagram (with optional code EM)

5) Enter the measured value manually.
6) “Result” is displayed automatically.

F060314.ai

Figure 6.3.14 “Active Pulse Output” Result