6 operation, A) axfa11, Figure 6.3.14 “active pulse output” result – Yokogawa FieldMate User Manual
Page 65
<6 OPERATION>
6-33
IM 01R01A11-01EN
NOTE
• Measure for more than 60 seconds.
• If using a YOKOGAWA CA150, set the mode to “Frequency (FREQ) and Pulse” and the
range to “CPM”.
• An appropriate load (24 VDC ± 20%, 150 mA max) is required.
4) Measure the pulse output value in “pps” with a Yokogawa CA150 Handy Calibrator or
equivalent to simplify the measurement of pulses. If such a device is unavailable, use the
confi guration shown in
Figure 6.3.13.
SB
B
A
SA
C
AL–
AL+
COM
SO2+
SO1+
I+
I–
CUR OUT
STATUS OUT
ALARM OUT
SIGNAL
COM
SI2+
SI1+
EX2
EX1
L/+
N/–
POWER SUPPLY
EXCITER
P–
P+
PULSE OUT
STATUS IN
Electronic Counter
Output voltage: 24 V DC ±20%
Current: 150 mA max
Load
Dashed box : Yokogawa CA150 Handy Calibrator or similar
F060313-A.ai
(A) AXFA11
Electronic Counter
Output voltage: 24 V DC ±20%
Current: 150 mA max
Load
Dashed box : Yokogawa CA150 Handy Calibrator or similar
F060313-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 6.3.13
Active Pulse Output Diagram (with optional code EM)
5) Enter the measured value manually.
6) “Result” is displayed automatically.
F060314.ai
Figure 6.3.14 “Active Pulse Output” Result