1 package and installation, 2 axf write protect cancellation, 1 before verification – Yokogawa FieldMate User Manual
Page 14: Package and installation -2, Axf write protect cancellation -2 3.2.1, Before verification -2
<3 PREPARATION>
3-2
IM 01R01A11-01EN
3.1
Package and Installation
Regarding the packaging and installation procedure, refer to IM 01R01A01-01E Part B.
3.2
AXF Write Protect Cancellation
3.2.1
Before Verification
1) AXF parameters are changed during the verification process. Cancel the write protect
switch.
Switch 1
Switch 2
Enable Protect
2 Burnout setting switch
1 Write protect setting switch
Hardware switch off
Set the "Write Protect Switch" to "Enable" before this verification.
Return the switch to the original position after this verification is
complete.
F030201-A.ai
(A) AXFA11
Hardware switch off
Set the "Write Protect Switch" to "Enable" before this verification.
Return the switch to the original position after this verification is
complete.
Switch 1
Switch 2
Enable Protect
2 Burnout setting switch
1 Write protect setting switch
F030201-B.ai
(B) AXFA14 or AXF Integral Flowmeter
Figure 3.2.1 Write Protect Cancellation
2) After the verification is complete, return the switch to the original position.