<6 OPERATION>

6-12

IM 01R01A11-01EN

6.2.1

“Circuit” check and “Device Status” check

(Verification Tool is in “online” mode)

CAUTION

• “Circuit” check and “Device Status” check temporarily change the parameters of AXF. In

case of unexpected termination, follow 8.2.

• The AXF flowtube should be completely filled with fluid, otherwise the results may be

inadequate.

• These items are run only when Verification Tool is in “online” mode.
• These items are verified for 2 types. Each condition check can be run separately.

(A) “No Flow” condition:

Verifies the condition of the AXF by checking;

- Circuits including the excitation circuit and the magnetic circuit inside the flowtube when

the fluid velocity is completely zero.

- The present and past history of alarms.

(B) “Flow” condition:

Verifies the condition of the AXF by checking;
- Circuits including the excitation circuit and the operation circuit which calculates the flow

velocity when the fluid velocity is sufficient.

- The present and past history of alarms.
• There are 3 combinations:

(1) No Flow
(2) Flow
(3) Flow + No Flow

• The check items of the “Circuit” check depend on the case as follows:

• Nominal size of the unit.
• With or without the optional code ELC (DC Noise Cut Circuit).
• Test condition type; “Flow” or “No Flow”.

For details, see

Table 6.2.2.