A) no flow (the fluid velocity is completely zero) – Yokogawa FieldMate User Manual
Page 47
<6 OPERATION>
6-15
IM 01R01A11-01EN
(A) No Flow (The fluid velocity is completely zero)
1) Fill the flowtube with fluid completely. Be sure that the fluid velocity is completely zero.
2
3
1
F060203.ai
Figure 6.2.3 “Flow/ No Flow” Screen
Table 6.2.3 “Flow/ No Flow” Screen
Item
No.
Item Name
Description
1
No Flow
Selects the “No Flow” test condition.
2
Start Verification (*1)
Runs “Circuit” check and “Device Status” check.
These checks are performed automatically. The Progress Bar is displayed
at the bottom left.
3
Cancel
Stops running “Circuit” check and “Device Status” check.
*1:
When the AXF protect function is set to “Protect”,
Figure 3.2.2 is displayed. Follow 3.2.2.
2) Select “No Flow (1)” and click “Start Verification (2)”.
3) Before running “Circuit” check and “Device Status” check,
Figure 6.2.4 is displayed.
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)