beautypg.com

Chapter 1 maintenance tests, Propaq md overview – ZOLL Propaq MD Rev A User Manual

Page 9

background image

9650-0807-01 Rev. A

Propaq MD Service Manual

5

Chapter 1

Maintenance Tests

Propaq MD Overview

This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a
trained biomedical technician must perform a CO

2

calibration and an NIBP calibration check

once per year. A CO

2

calibration check may be performed at any time to verify the CO

2

calibration. It is necessary to perform maintenance tests after a module is repaired to ensure that
the functions of the Propaq MD unit work properly. Use the checklist at the back of this
document (ZOLL Propaq MD Maintenance Tests Checklist) to record your results of the
maintenance tests.

Because the Propaq MD units must be maintained ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL Propaq MD Operator’s Guide for the Operator’s Shift Checklist.)