beautypg.com

Shock test – ZOLL Propaq MD Rev A User Manual

Page 26

background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

22

9650-0807-01 Rev. A

Shock Test

WARNING!

Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.

Keep hands and all other objects clear of the multi-function cable connections and
defibrillator analyzer when discharging the defibrillator.

Before you discharge the defibrillator, warn everyone near the equipment to STAND
CLEAR.

Equipment

Fluke Impulse 7000 DP Analyzer (or equivalent), stop watch

Do This...

Observe This...

Pass/Fail

1.

Turn on the Analyzer.

2.

On the Analyzer, select the DEFIB function.

3.

On the Analyzer, select ENERGY.

4.

Connect the Propaq MD MFC connector to
the pads connection.

5.

On the Propaq MD unit, apply power to the
unit using a battery or an AC adapter.

6.

Connect the MFC cable to the Propaq MD
unit.

7.

Turn on the Propaq MD unit, and press the

Leads quick access key

(

)

to select

PADS

.

8.

On the Propaq MD unit, set energy level to
5J by pressing the Energy Select DOWN
arrow.

9.

On the Propaq MD unit, press the
CHARGE button.

The unit charges to 5J, the red
Shock button illuminates and a
ready tone sounds.

o

o

10.

On the Propaq MD unit, press the Shock
button.

11.

On the Analyzer, observe and record the
energy measurement.

Energy is delivered at 5J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 25.

o

o

12.

On the Propaq MD unit, set the energy level
to 50J by pressing the Energy Select UP
arrow.

I, II,

III...