ZOLL Propaq MD Rev A User Manual
Page 61
Propaq MD Error Messages
9650-0807-01 Rev. A
Propaq MD Service Manual
57
TEMP DEVICE FAILURE
Temp software detected a
problem with the
temperature data that is
most likely a hardware issue.
Replace the main board.
TEMP PROBE1 (or
PROBE2) SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.
TRANSMISSION FAILED
The communication
operation failed. This
message displays until a
new operation is started.
To correct the problem,
verify the following:
•
Wireless communication
is enabled on the
Propaq MD unit.
•
The Wi-Fi settings are
correct in the
Communications setup
menu.
•
The ZOLL server is
configured correctly.
•
Your cellular device is
configured correctly.
•
The Propaq MD unit is
within range of the
wireless server.
WAN CONNECTION
UNAVAILABLE
The Propaq MD unit was
unable to establish a WAN
connection.
To correct the problem,
verify the following:
•
Wireless communication
is enabled on the
Propaq MD unit.
•
The expected means of
coummunication is
enabled (for instance, if
using a cell modem,
make sure the USB cell
modem is set to
Enabled).
•
(Bluetooth) It is paired
with the Propaq MD unit.
•
(Wi-Fi) The correct Wi-Fi
access point profile is
selected.
•
(Cellular modem) It is
properly connected to
the Propaq MD unit and
that the USB icon on the
display is illuminated.
Error Message
Explanation
Technical Action