beautypg.com

Temperature test – ZOLL Propaq MD Rev A User Manual

Page 47

background image

Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

43

Temperature Test

Equipment

IBP Temperature Simulator (or equivalent)

Do this...

Observe This...

Pass/Fail

1.

Connect the temperature simulator to the
Propaq MD T1 channel.

2.

On the IBP Temp Simulator, Select 98.6
Degree F (3

7 Degree C) or a setting as close

as possible as your simulator allows.

The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.

o

o

3.

Repeat steps 1 and 2 of this procedure for the
T2 channel.

4.

Disconnect cable.