Propaq MD Error Messages

9650-0807-01 Rev. A

Propaq MD Service Manual

53

NIBP AIR LEAK

Major air leak or could not
reach target pressure.

Attempt to identify
problems with the hose
and/or cuff by doing the
following:

Check the hose
connection to the
device. Make sure that
the hose is tightly
screwed onto the NIBP
connector.

Ensure that the cuff is
appropriately wrapped
and is not too loose.

Inspect the cuff and
hose for leaks. Inspect
the O-ring on the NIBP
hose connector.
Consider replacing the
cuff and/or hose.

If the problem persists,
perform the following
procedure: “NIBP Leak
Test” on page 39
:
Inspect internal NIBP
tubing for leaks.

NIBP ARTIFACT

Unable to determine blood
pressure. Check
connections; restrict patient
movement.

Attempt to identify and
eliminate the source of
artifact. If problem persists,
replace the NIBP module.

NIBP CALIBRATION
CHECK DUE

Annual calibration check is
due.

Perform the “NIBP
Calibration Check” on
page 36
.

NIBP COMMUNICATION
FAILURE

There is a communication
problem with the NIBP
module.

Ensure that the J32
connection is secure. If
error persists, replace the
NIBP module.

NIBP EQUIPMENT CHECK

Device is unable to obtain a
reading.

Check the hose and cuff
connections. If problem
persists, perform one or
more of the following
procedures in Chapter 1:

NIBP Leak Test

NIBP Deflation Test

NIBP Inflation Test

Error Message

Explanation

Technical Action